Characteristics:
Multiparameter meter contains 15 optional parameters pH, ORP, Ion, conductivity, TDS, salinity, resistivity and DO of the samples.
pH Mode:
• 1 to 5 points push-button calibration with auto-buffer recognition.
• Automatic electrode slope display helps user decide whether to replace the pH electrode.
ORP Mode:
• Single point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt measurements provide accurate ORP readings.
Ion Mode:
• 2 to 5 points calibration including the eight concentration calibration points can be selected.
• Selectable multiple concentration units including the ppm, mg/L and mol/L.
Conductivity/TDS/Salinity/Resistivity Mode:
• 1 to 5 points push-button calibration, automatically recognizes the standard solutions.
• Selectable cell constant, normalization temperature, TDS conversion factor, seawater and
practical salinity measurement modes.
Dissolved Oxygen Mode:
• 1 or 2 points calibration using the air-saturated water and zero oxygen solution.
• Manual salinity and barometric pressure compensations ensure accurate measureme
Country Of Origin : India
pH Range: -2.000~20.000pH
Accuracy: ±0.002pH
Resolution: 0.1, 0.01, 0.001pH, Selectable
Calibration Points: 1 to 5 points
Calibration Solutions: USA, NIST, DIN or User-defined
Temperature Compensation: 0~100°C, 32~212°F
Compensation Mode: Manual or Automatic
mV Range: -1999.9~1999.9mV
Accuracy: ±0.2mV
Resolution: 0.1, 1mV, Selectable
Calibration Points: 1 point (Only for Relative mV mode)
Calibration Range: ±200mV
Ion Range: 0.001~19999ppm, mg/L, mol/L (Refer to the range of ISE)
Accuracy: ±0.5% F.S (Monovalent), ±1% F.S (Divalent)
Resolution: 0.001, 0.01, 0.1, 1
Calibration Points: 2 to 5 points
Calibration Solutions: 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mg/L, mol/L
Temperature Compensation: 0~100°C, 32~212°F
Compensation Mode: Manual or Automatic
Conductivity Range: 0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
Accuracy: ±0.5% F.S
Calibration Points: 1 to 5 points
Calibration Solutions: 10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm
Temperature Compensation: 0~100°C, 32~212°F, Manual or Automatic
Temperature Coefficient: 0.0~10.0%/°C
Compensation Modes: Linear or Pure Water
Cell Constant: K=0.1, 1, 10 or User-defined
Normalization Temperature: 20°C or 25°C
TDS Range: 0~100ppt (Max. 200ppt, depending on factor setting)
Accuracy: ±1% F.S
TDS Factor: 0.1~1.0 (Default 0.5)
Salinity Range: 0~10ppt (Max. 80ppt)
Accuracy: ±1% F.S
Measurement Modes: Seawater or Practical Salinity
Resistivity Range: 0~100MΩ
Accuracy: ±1% F.S
Resolution: 0.01, 0.1, 1
Dissolved Oxygen Range: 0.00~20.00mg/L (or ppm)
Accuracy: ±0.2mg/L
Resolution: 0.01mg/L
Calibration Points: 1 or 2 points
Temperature Compensation: 0~50°C, 32~122°F, Automatic
Pressure Correction: 60.0~112.5kPa, 450~850mmHg
Salinity Correction: 0~50g/L
% Saturation of Oxygen Range: 0.0~200.0%
Accuracy: ±2.0%
Resolution: 0.1, 1%
Temperature Range: 0~105°C, 32~221°F
Accuracy: ±0.5°C, ±0.9°F
Resolution: 0.1°C
Calibration Points: 1 point
Calibration Range: Measured value ±10°C
Others Features: